AP-103110-970

INCUBATOR

• Touch switch panel with LED digital display, set-temp, real -temp are displayed separately; multiply self-checking alarms with audible and visible

• Air, skin humidity and oxygen sensor

• 6 openings to access the baby

• Heating power rate 0-100%

• Isolated alarm in case of failure of the alarm system

• Inclination of the bassinet: ±12°

• Double isolation against heat loss

Standard configuration:

• Double wall hood

• Infant bed with mattress

• Temperature controller

• Water tank

• Air filter

• Blue light radiant box

• I.V. Pole, stand and shelf

• Oxygen concentration control system 


• Dimensions: 106cm × 93cm × 59cm

• Volume: 1.19m³

• Weight : 60Kg

Contact us

  • 16

    16 years of experience

  • Large distribution network

  • Quality technical services

More info

Technical specification:

• Power supply: AC220V-50HZ±

• Power input: 600VA

• Air T° control range: 25°C~37°C

• Changing of temperature: ≤0.5°C

• Uniformity of temperature: ≤0.8°C

• Temperature rising time: ≤45min

• Noise level: ≤ 55dB( A)

Temperature control ranges:

• Air mode control: 25°C~37°C and 37°C ~ 39°C

• Skin mode control: 34°C~37°C et 37°C~38°C

• Temperature rise time: ≤35min

• Temperature variability: ≤0.5°C

• Temperature uniformity: ≤0.8°C

• Precision of skin sensor: <0.3°C

Oxygen Control: 

• Oxygen display range: 18% ~ 100%

• Oxygen display resolution: 1%

• Oxygen display accuracy: ± 3%

• Oxygen set range: 20% ~ 60%

• Life of oxygen sensor: at most 10.000 hours at 100% oxygen.

Alarms:

• Audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater

• Over-temp alarm: when ≤38°C

• Humidity alarm: low or high humidity

• Oxygen alarm: low or high oxygen

• Temperature alarm: high/low air or skin t°

• Sensor alarm when the sensor is disconnected, when the circuit is open, electricity problem, ventilator is blocked, does not work or runs at a speed, is lower than 800r/m or does not work

• Air temperature sensor failure

• Skin temperature failure

• Isolate overheat sensor failure

• Air flow sensor failure

• Power failure alarm: When no power or is disconnected 

• Motor alarm: Motor stop running or the speed is lower than 800 r/m

• Alarm Water reservoir is empty or placed incorrectly 